Safety Considerations of 7-T MRI in Clinical Practice

Radiology. 2019 Sep;292(3):509-518. doi: 10.1148/radiol.2019182742. Epub 2019 Jul 16.

Abstract

Although 7-T MRI has recently received approval for use in clinical patient care, there are distinct safety issues associated with this relatively high magnetic field. Forces on metallic implants and radiofrequency power deposition and heating are safety considerations at 7 T. Patient bioeffects such as vertigo, dizziness, false feelings of motion, nausea, nystagmus, magnetophosphenes, and electrogustatory effects are more common and potentially more pronounced at 7 T than at lower field strengths. Herein the authors review safety issues associated with 7-T MRI. The rationale for safety concerns at this field strength are discussed as well as potential approaches to mitigate risk to patients and health care professionals.

Publication types

  • Review

MeSH terms

  • Equipment Safety / methods*
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Risk Assessment