Initial Clinical Experience Treating Patients with Breast Cancer on a 6-MV Flattening-Filter-Free O-Ring Linear Accelerator

Adv Radiat Oncol. 2019 Jun 8;4(4):571-578. doi: 10.1016/j.adro.2019.05.006. eCollection 2019 Oct-Dec.

Abstract

Purpose: Three-dimensional (3D) conformal radiation therapy is the standard technique used for adjuvant breast radiation. We report the clinical use of a novel 6-MV flattening-filter-free O-ring linear accelerator (6X-FFF ORL) for breast cancer that may improve upon 3D conformal radiation therapy with its higher dose rate and faster rotation and leaf speed than traditional C-arm gantries.

Methods and materials: We retrospectively identified consecutive women with breast cancer who underwent surgery followed by radiation therapy to the breast or chest wall on Halcyon (Varian Medical Systems, Palo Alto, CA), a novel 6X-FFF ORL. We report their clinicopathologic information, radiation therapy details, acute toxicities, dose-volume histogram data, couch corrections, and treatment times.

Results: Thirty-four women were treated for breast cancer on a 6X-FFF ORL between February 2018 and September 2018. All patients underwent lumpectomy (92%) or mastectomy (8%). Tumors were left sided in 44% and bilateral in 9%, and 9% included comprehensive nodal radiation therapy. Twelve percent of patients were treated prone and 29% with deep-inspiration breath hold. Standard target and normal-tissue constraints were met in nearly all plans. The 3D vector couch correction average was 0.77 ± 0.05 cm. The mean beam-on time was 2.0 ± 0.3 minutes, and mean treatment time from start of imaging to beam-off was 4.4 ± 0.4 minutes. Grade 2 dermatitis, fatigue, and breast pain occurred in 18%, 9%, and 3% of patients, respectively.

Conclusions: In this first clinical report of breast radiation therapy with a 6X-FFF ORL, treatment was versatile and fast for complex setups and techniques, with acceptable toxicity and organ-at-risk doses. Thus, a 6X-FFF ORL can increase throughput or reduce length of day compared with a conventional C-arm linear accelerator in departments with a busy breast service.