Accuracy and dimensional reproducibility by model scanning, intraoral scanning, and CBCT imaging for digital implant dentistry

Int J Implant Dent. 2021 Jun 30;7(1):63. doi: 10.1186/s40729-021-00343-w.

Abstract

Background: Recently, it has become possible to analyze implant placement position using the digital matching data of optical impression data of the oral cavity or plaster models with cone beam computed tomography (CBCT) data, and create a highly accurate surgical guide. It has been reported that CBCT measurements were smaller than the actual values, termed shrinkage. Matching of digital data is reliable when the plaster model or intraoral impression values show shrinkage at the same rate as the CBCT data. However, if the shrinkage rate is significantly different, the obtained digital data become unreliable. To clarify digital matching reliability, we examined dimensional reproducibility and shrinkage in measurements obtained with a model scanner, intra-oral scanner (iOS), and CBCT.

Materials and methods: Three implants that were arranged in a triangle were fixed in an acrylic plate. The distance between each implants were measured using model scanner, iOS, and CBCT. The actual size measured by electronic caliper was regarded as control.

Results: All values measured with CBCT were significantly smaller than that of model scanner, iOS, and control (p<0.001). The model scanner shrinkage was 0.37-0.39%, iOS shrinkage was 0.9-1.4%, and CBCT shrinkage was 1.8-6.9%. There were statistically significant differences among the shrinkage with iOS, CBCT, and model scanner (p<0.001).

Conclusion: Our findings showed that all measurements obtained with those modalities showed shrinkage as compared to the actual values. In addition, CBCT shrinkage was largest among three different measuring methods. They indicated that data matching between CBCT and scanner measurements requires attention in regard to the reliability of values obtained with those devices.

Keywords: CBCT; Digital implant dentistry; Intraoral scanner; Model scanner; Surgical guide.

MeSH terms

  • Cone-Beam Computed Tomography
  • Dentistry
  • Imaging, Three-Dimensional
  • Reproducibility of Results
  • Spiral Cone-Beam Computed Tomography*