Retropharyngeal Reduction Plate for Atlantoaxial Dislocation: A Cadaveric Test and Morphometric Trajectory Analysis

Orthop Surg. 2022 Mar;14(3):522-529. doi: 10.1111/os.13217. Epub 2022 Jan 30.

Abstract

Objective: To evaluate the placement feasibility and safety of the newly designed retropharyngeal reduction plate by cadaveric test and to perform morphometric trajectory analysis.

Methods: The five cadaveric specimens with intact atlantoaxial joint were enrolled in this study. They were used for simulating the placement process and evaluating the placement feasibility of the retropharyngeal reduction plate. The atlantoaxial dislocation (AAD) of five cadaveric specimens were obtained by proper external force after dissecting ligaments. The retropharyngeal reduction plate was placed on atlantoaxial joint of cadaveric specimens. The X-ray and three-dimensional (3D) spiral CT were used for evaluating the placement safety of retropharyngeal reduction plate. The DICOM data was obtained after 3D spiral CT scanning for the morphometric trajectory analysis.

Results: The reduction plates were successfully placed on the atlantoaxial joint of five cadaveric specimens through the retropharyngeal approach, respectively. The X-ray and 3D spiral CT showed the accurate screw implantation and satisfying plate placement. The length of the left/right atlas screw trajectory (L/RAT) was, respectively, 1.73 ± 0.01 cm (LAT) and 1.71 ± 0.02 cm (RAT). The length of odontoid screw trajectory (OST) was 1.38 ± 0.02 cm. The length of the left/right axis screw trajectory (L/RAXT) was, respectively, 1.67 ± 0.02 cm (LAXT) and 1.67 ± 0.01 cm (RAXT). There was no statistical significance between left side and right side in terms of AT and AXT (P > 0.05). The angles of atlas screw trajectory angle (ASTA), axis screw trajectory angle (AXSTA), and odontoid screw trajectory angle (OSTA) were 38.04° ± 2.03°, 56.92° ± 2.66°, and 34.78° ± 2.87°, respectively.

Conclusion: The cadaveric test showed that the retropharyngeal reduction plate is feasible to place on the atlantoaxial joint, which is also a safe treatment choice for atlantoaxial dislocation. The meticulous preoperative planning of screw trajectory based on individual differences was also vital to using this technique.

Keywords: Atlantoaxial dislocation; Cadaveric study; Craniovertebral junction; Morphometric trajectory analysis; Retropharyngeal approach.