Accuracy of intravitreal injection volume for aflibercept pre-filled syringe and BD Luer-Lok one-milliliter syringe

Int J Retina Vitreous. 2022 Apr 5;8(1):27. doi: 10.1186/s40942-022-00375-3.

Abstract

Background: To evaluate the accuracy of intravitreal injection volume of the pre-filled syringe (PFS) in which aflibercept is packaged compared to the BD Luer-Lok 1-mL syringe.

Methods: Ophthalmologists injected their typical intravitreal volume for aflibercept using either the PFS or BD Luer-Lok 1-mL syringe for 5 times each. The injected fluid was weighed using a micro-scale and converted to volume. The volume of fluid injected was also evaluated when the 0.05 mL line on the PFS was lined up to the tip or base of the dome-shaped plunger.

Results: Injection volume was measured for 12 physicians. The average injected fluid volume was 74.22 ± 15.87 µL for PFS and 53.42 ± 4.61 µL for the BD Luer-Lok 1-mL syringe (p < 0.0001). The average deviation in volume injected for the PFS was higher compared to the BD Luer-Lok 1-mL syringe (11.36 µL vs. 3.35 µL, p < 0.0001). When the PFS was lined up with the tip of the dome-shaped plunger at the 0.05-mL line, the average injected volume was 71.03% higher.

Conclusions: The intravitreal injection volume and variability using the new PFS were significantly higher than the volume injected using the BD Luer-Lok 1-mL syringe previously used, potentially leading to higher rates of visually significant elevation of intraocular pressures.

Keywords: Aflibercept; Intravitreal injection; Pre-filled syringe.