A point-of-need platform for rapid measurement of a host-protein score that differentiates bacterial from viral infection: Analytical evaluation

Clin Biochem. 2023 Jul:117:39-47. doi: 10.1016/j.clinbiochem.2022.04.012. Epub 2022 Apr 26.

Abstract

The objective was to evaluate the analytical performance of a new point-of-need platform for rapid and accurate measurement of a host-protein score that differentiates between bacterial and viral infection. The system comprises a dedicated test cartridge (MeMed BV®) and an analyzer (MeMed Key®). In each run, three host proteins (TRAIL, IP-10 and CRP) are measured quantitatively and a combinational score (0-100) computed that indicates the likelihood of Bacterial versus Viral infection (BV score). Serum samples collected from patients with acute infection representing viral (0 ≤ score < 35), equivocal (35 ≤ score ≤ 65), or bacterial (65 < score ≤ 100) scores based on pre-defined score cutoffs were employed for the analytical evaluation studies as well as samples from healthy individuals. To assess reproducibility, triplicate runs were conducted at 3 different sites, on 2 analyzers per site over 5 non-consecutive days. Lower limit of quantitation (LLoQ) and analytical measurement range were established utilizing recombinant proteins. Sample stability was evaluated using patient samples representative of BV score range (0-100). MeMed Key® and MeMed BV® passed the acceptance criteria for each study. In the reproducibility study, TRAIL, IP-10 and CRP measurements ranged with coefficient of variation from 9.7 to 12.7%, 4.6 to 6.2% and 5.0 to 11.6%, respectively. LLoQ concentrations were established as 15 pg/mL, 100 pg/mL and 1 mg/L for TRAIL, IP-10 and CRP, respectively. In summary, the analytical performance reported here, along with diagnostic accuracy established in the Apollo clinical validation study (NCT04690569), supports that MeMed BV® run on MeMed Key® can serve as a tool to assist clinicians in differentiating between bacterial and viral infection.

Keywords: CRP; Chemiluminescence; IP-10; MeMed BV; MeMed Key; Rapid immunoassay platform; TRAIL.

MeSH terms

  • C-Reactive Protein*
  • Chemokine CXCL10
  • Humans
  • Reproducibility of Results
  • Virus Diseases* / diagnosis

Substances

  • C-Reactive Protein
  • Chemokine CXCL10