Purpose: This study aims to evaluate the performance of low-cost homemade spacers compared with manufactured valved holding chambers (VHCs) for fluticasone propionate delivery via a pMDI (pressurized Metered Dose Inhaler).
Methods: The Total Emitted Dose (TED) and particle size distribution were measured for pMDI alone or connected to the different spacers, according to CAN/CSA-Z264.1-02 standard. Two types of low-cost alternative and manufactured spacers were investigated: 500 mL plastic bottle and 553 mL aluminium can; non-antistatic plastic VHCs and aluminium antistatic VHCs.
Results: The TED of homemade plastic bottle vs plastic VHC were similar in the 20-23% range. In contrast, the TED of homemade aluminium can was higher compared to aluminium VHC (83% vs 68%). The Fine Particle Fraction (FPF) was similar for the two plastic-based spacers (in the 12.68-17.60% range), although it was greater for the aluminium can compared to aluminium VHC (51% vs 42%). However, all spacers have limited large particles fraction, mainly deposited in the oropharyngeal tract, potentially decreasing side effects.
Conclusion: We demonstrated that low-tech solutions as homemade spacers have at least similar performances to VHC medical devices composed of same material (aluminium or plastic). Thus, low-cost homemade spacers represent alternatives in case of emergency and without VHCs nearby.
Keywords: Aerosol delivery; Cost effective; Developing countries; Low-tech; Spacer; Valved holding chamber.
Copyright © 2023 The Author(s). Published by Elsevier B.V. All rights reserved.