A carbon nanotube x-ray source array designed for a new multisource cone beam computed tomography scanner

Phys Med Biol. 2024 Mar 26;69(7). doi: 10.1088/1361-6560/ad3323.

Abstract

Cone beam computed tomography (CBCT) is known to suffer from strong scatter and cone beam artifacts. The purpose of this study is to develop and characterize a rapidly scanning carbon nanotube (CNT) field emission x-ray source array to enable a multisource CBCT (ms-CBCT) image acquisition scheme which has been demonstrated to overcome these limitations. A CNT x-ray source array with eight evenly spaced focal spots was designed and fabricated for a medium field of view ms-CBCT for maxillofacial imaging. An external multisource collimator was used to confine the radiation from each focal spot to a narrow cone angle. For ms-CBCT imaging, the array was placed in the axial direction and rapidly scanned while rotating continuously around the object with a flat panel detector. The x-ray beam profile, temporal and spatial resolutions, energy and dose rate were characterized and evaluated for maxillofacial imaging. The CNT x-ray source array achieved a consistent focal spot size of 1.10 ± 0.04 mm × 0.84 ± 0.03 mm and individual beam cone angle of 2.4°±0.08 after collimation. The x-ray beams were rapidly switched with a rising and damping times of 0.21 ms and 0.19 ms, respectively. Under the designed operating condition of 110 kVp and 15 mA, a dose rate of 8245μGy s-1was obtained at the detector surface with the inherent Al filtration and 2312μGy s-1with an additional 0.3 mm Cu filter. There was negligible change of the x-ray dose rate over many operating cycles. A ms-CBCT scan of an adult head phantom was completed in 14.4 s total exposure time for the imaging dose in the range of that of a clinical CBCT scanner. A spatially distributed CNT x-ray source array was designed and fabricated. It has enabled a new multisource CBCT to overcome some of the main inherent limitations of the conventional CBCT.

Keywords: CBCT; carbon nanotube; cone beam artifacts; field emission x-ray; multisource; scatter reduction; x-ray imaging.

MeSH terms

  • Cone-Beam Computed Tomography / methods
  • Nanotubes, Carbon*
  • Phantoms, Imaging
  • Tomography, X-Ray Computed / methods
  • X-Rays

Substances

  • Nanotubes, Carbon