Discrimination of Multiple Homologous Sequences Based on DNA Logic Gate and Elemental Labeling Technology

Anal Chem. 2024 Apr 23;96(16):6329-6336. doi: 10.1021/acs.analchem.3c05915. Epub 2024 Apr 10.

Abstract

The simultaneous discrimination of multiple homologous sequences faces challenges due to the high similarity of sequences and the complexity of the discrimination system in most reported works. Herein, a simple and ingenious analysis method was developed to identify eight miRNAs of the let-7 family by combining logic gates and entropy-driven catalytic (EDC)-based lanthanide labeling inductively coupled plasma mass spectrometry (ICP-MS) technology. Specifically, eight miRNAs were first divided into four types according to the difference of bases in the domains 2 and 3 on sequences. To identify the type of targets, a DNA logic gate was constructed with two strand displacement reactions on magnetic beads that could be initiated by different types of targets. Based on the difference of the output signals after two strand displacement reactions, the type of targets was distinguished preliminarily. Then, the discrimination of a specific target was achieved with EDC-based lanthanide labeling ICP-MS detection. By labeling the different magnetic probes with different elemental tags, a specific element signal released from magnetic beads after EDC could be detected by ICP-MS, and therefore, simultaneous detection of homologous sequences was completed. This work provided a novel and simple method for highly specific identification of homologous sequences with the assistance of a logic gate and can promote further development of elemental labeling ICP-MS in the field of multiple analysis.