Feasibility and safety report on robotic assistance for cervical pedicle screw fixation: a cadaveric study

Sci Rep. 2024 May 13;14(1):10881. doi: 10.1038/s41598-024-60435-6.

Abstract

This cadaveric study aimed to evaluate the safety and usability of a novel robotic system for posterior cervical pedicle screw fixation. Three human cadaveric specimens and C2-T3 were included. Freshly frozen human cadaver specimens were prepared and subjected to robot-assisted posterior cervical pedicle screw fixation using the robotic system. The accuracy of screw placement, breach rate, and critical structure violations were evaluated. The results were statistically compared with those of previous studies that used different robotic systems for cervical pedicle screw fixation. The robotic system demonstrated a high accuracy rate in screw placement. A significant number of screws were placed within predetermined safe zones. The total entry offset was 1.08 ± 0.83 mm, the target offset was 1.86 ± 0.50 mm, and the angle offset was 2.14 ± 0.77°. Accuracy rates comparable with those of previous studies using different robotic systems were achieved. The system was also feasible, allowing precise navigation and real-time feedback during the procedure. This cadaveric study validated the safety and usability of the novel robotic system for posterior cervical pedicle screw fixation. The system exhibited high precision in screw placement, and the results support the extension of the indications for robot-assisted pedicle screw fixation from the lumbar spine to the cervical spine.

MeSH terms

  • Cadaver*
  • Cervical Vertebrae* / surgery
  • Feasibility Studies*
  • Female
  • Humans
  • Male
  • Pedicle Screws*
  • Robotic Surgical Procedures* / instrumentation
  • Robotic Surgical Procedures* / methods
  • Spinal Fusion / instrumentation
  • Spinal Fusion / methods