HemoQuant, a new quantitative assay for fecal hemoglobin. Comparison with Hemoccult

Ann Intern Med. 1984 Sep;101(3):297-302. doi: 10.7326/0003-4819-101-3-297.

Abstract

HemoQuant, an assay based on heme-derived porphyrin, quantifies both total fecal hemoglobin ant that fraction already converted to porphyrin by gut flora (the intestinal converted fraction). Recovery by HemoQuant of blood added to stools exceeded 99% and was not affected by ascorbic acid, iron, or other additives. HemoQuant was used to evaluate the performance of Hemoccult on 1018 stools. Hemoccult sensitivity differed widely, remaining negative with up to 42.5 mg hemoglobin per gram of stool (about 43 mL/d of blood) but positive with only 0.04 mg of hemoglobin per gram of stool. Driest stools tended to be Hemoccult-negative and the wettest Hemoccult-positive. Hemoccult-negative stools became positive after aqueous dilution independent of hemoglobin concentration. The intestinal converted fraction, not detected by guaiac tests, often constituted most of the fecal hemoglobin equivalents and was significantly higher in Hemoccult-negative than in Hemoccult-positive stools. Hemoccult lost reactivity during fecal storage due to progressive heme degradation. Frequent false-negative and false-positive reactions of Hemoccult were explained largely by variations in stool liquidity and heme degradation.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Diagnostic Errors
  • Evaluation Studies as Topic
  • Heme / metabolism
  • Hemoglobins / analysis*
  • Humans
  • Intestinal Mucosa / metabolism
  • Occult Blood*
  • Quality Control
  • Specimen Handling
  • Water