Simultaneous determination of cyclandelate and its metabolite in human plasma by capillary column gas-liquid chromatography

J Chromatogr. 1981 May 8;223(2):365-70. doi: 10.1016/s0378-4347(00)80109-7.

Abstract

A method was developed for the simultaneous determination of cyclandelate and mandelic acid concentrations in plasma, involving extraction from plasma followed by trimethylsilylation and chromatography of the derivatives on a glass capillary column with hydrogen flame-ionization detection. Calibration graphs were linear down to at least 20 microgram/ml for each substance. The precision was excellent with a pooled relative standard deviation of 6.3% and 6.4% for cyclandelate and mandelic acid serum samples, respectively. Concentrations below 500 ng/ml of each substance could be detected in human plasma. The method was developed for use in bioavailability and metabolism studies.

MeSH terms

  • Animals
  • Chromatography, Gas
  • Cyclandelate / blood*
  • Humans
  • Mandelic Acids / blood*
  • Reference Values

Substances

  • Mandelic Acids
  • Cyclandelate