The releasable balloon technique with activated high frequency electrical current

Surg Neurol. 1980 Jun;13(6):405-8.

Abstract

A new releasable balloon catheter was developed which has a soluble joint connecting the balloon and the catheter. The joint is made from a polyvinyl alcohol tube and has two electrodes. When a high frequency electrical current is activated, the resulting heat causes the joint to disappear within one second. The type of generator used in the experiment was an Aesculap GK-34. This catheter was used for embolization of carotid-jugular fistulas in dogs.

MeSH terms

  • Animals
  • Arteriovenous Fistula / surgery
  • Carotid Artery Diseases / surgery
  • Catheterization / instrumentation
  • Catheterization / methods*
  • Cerebrovascular Disorders / surgery*
  • Dogs
  • Electricity
  • Embolization, Therapeutic / instrumentation
  • Embolization, Therapeutic / methods*
  • Jugular Veins / surgery