Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography

Radiology. 1996 Jan;198(1):55-60. doi: 10.1148/radiology.198.1.8539406.

Abstract

Purpose: To test the hypothesis that respiration effects in three-dimensional (3D) coronary magnetic resonance (MR) imaging can be reduced with navigator-echo-based gating or triggering according to the superior-inferior position of the diaphragm.

Materials and methods: Real-time respiratory gating and respiratory triggering (breath hold with feedback) were implemented with navigator echoes in a magnetization-prepared, segmented, 3D coronary imaging sequence. The two techniques were first tested with a motion phantom. An imaging protocol that compared real-time respiratory-gated acquisition, real-time respiratory-triggered acquisition, and continuous acquisition was then evaluated in six healthy subjects.

Results: Real-time respiratory-gated and respiratory-triggered acquisition were superior to continuous acquisition with two signals averaged (P = .025). The performance of the gated acquisition was about the same as that of the triggered acquisition (P = .05).

Conclusion: Navigator-echo-based, real-time respiratory-gating and respiratory-triggering techniques are practical methods for effective reduction of respiration effects in coronary MR imaging.

Publication types

  • Comparative Study
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Artifacts
  • Coronary Vessels / anatomy & histology*
  • Electrocardiography
  • Heart Rate
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods
  • Magnetic Resonance Angiography / methods*
  • Middle Aged
  • Phantoms, Imaging
  • Respiration*