Abstract

This paper describes the use of nonaqueous capillary electrophoresis (NACE) with laser induced fluorescence (LIF) to improve detection sensitivity. The nonaqueous medium is conducive to lower detection limits due to the minimization of quenching effects. The nonaqueous solvent, N-methylformamide, produced the best detection limit with a 2-fold enhancement using NACE-LIF as compared to aqueous CE. N-methylformamide also gave the best fluorescent signal enhancement (6-fold) among the five nonaqueous solvents tested using steady-state fluorescence. The extent (or degree) of enhancement in fluorescence intensity seems to be related to the viscosity and/or polarity of the solvent.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Calibration
  • Electrophoresis, Capillary / methods*
  • Fluorescence
  • Formamides
  • Lasers*
  • Naphthalenes
  • Sensitivity and Specificity
  • Solvents

Substances

  • Formamides
  • Naphthalenes
  • Solvents
  • 8-amino-1,3,6-naphthalenetrisulfonic acid
  • methylformamide