Optic nerve sheath diameter (ONSD) ultrasound is becoming increasingly more popular for estimating raised intracranial pressure (ICP). We performed a systematic review and analysis of the diagnostic accuracy of ONSD when compared to the standard invasive ICP measurement.
Method: We performed a systematic search of PUBMED and EMBASE for studies including adult patients with suspected elevated ICP and comparing sonographic ONSD measurement to a standard invasive method. Quality of studies was assessed using the QUADAS-2 tool by two independent authors. We used a bivariate model of random effects to summarize pooled sensitivity, specificity, and diagnostic odds ratio (DOR). Heterogeneity was investigated by meta-regression and sub-group analyses.
Results: We included 18 prospective studies (16 studies including 619 patients for primary outcome). Only one study was of low quality, and there was no apparent publication bias. Pooled sensitivity was 0.9 [95% confidence intervals (CI): 0.85-0.94], specificity was 0.85 (95% CI: 0.8-0.89), and DOR was 46.7 (95% CI: 26.2-83.2) with partial evidence of heterogeneity. The Area-Under-the-Curve of the summary Receiver-Operator-Curve was 0.93 (95% CI: 0.91-0.95, P < .05). No covariates were significant in the meta-regression. Subgroup analysis of severe traumatic brain injury and parenchymal ICP found no heterogeneity. ICP and ONSD had a correlation coefficient of 0.7 (95% CI: 0.63-0.76, P < .05).
Conclusion: ONSD is a useful adjunct in ICP evaluation but is currently not a replacement for invasive methods where they are feasible.
Keywords: intracranial pressure; meta-analysis; optic nerve sheath diameter; ultrasound.