Blood Recirculation Enhances Oxygenation Efficiency of Artificial Lungs.
Madhani SP, May AG, Frankowski BJ, Burgreen GW, Federspiel WJ.
Madhani SP, et al.
ASAIO J. 2020 May;66(5):565-570. doi: 10.1097/MAT.0000000000001030.
ASAIO J. 2020.
PMID: 31335366
Hydrodynamic performance and shear stresses of the device were analyzed using CFD at 2,000, 2,250 and 2,500 RPM. A prototype (0.38 m) was manufactured for in-vitro oxygenation testing. Oxygenation was measured at a constant 3.5 L/min blood flow while recirculation flow rat …
Hydrodynamic performance and shear stresses of the device were analyzed using CFD at 2,000, 2,250 and 2,500 RPM. A prototype (0.38 m) …