Micro-hole array sprayer combined with an organic membrane to assist LIBS (MASOM-LIBS): A novel highly sensitive detection method for dissolved trace heavy metals in water.
Shang J, Ling Y, Ping L, Pan K, Xue Y, Wang Y, Li J, Zhang S, Zhong S.
Shang J, et al.
Talanta. 2023 Nov 1;264:124780. doi: 10.1016/j.talanta.2023.124780. Epub 2023 Jun 7.
Talanta. 2023.
PMID: 37302350
The experimental results of Cu, Cd, Mn, Pb, Cr and Sr as target elements show that the LODs of the MASOM-LIBS method for most elements is less than 0.1 mg/L when the detection time is less than 3 min, which has certain advantages over similar LIBS methods. ...In view of th …
The experimental results of Cu, Cd, Mn, Pb, Cr and Sr as target elements show that the LODs of the MASOM-LIBS method for most element …